Differential Pressure – Piezoresistive Technical Data
-
Measuring range 0 to 300 hPa
-
Accuracy ±0.5 hPa (0.0 to +50.0 hPa)
-
±1 % of mv (+50.1 to +100.0 hPa)
-
±1.5 % of mv (Remaining Range)
-
Resolution 0.1 hPa
Flue gas O₂ Technical Data
-
Measuring range 0 to 21 Vol.%
-
Accuracy ±0.2 Vol.%
-
Resolution 0.1 Vol.%
CO determination (without H₂-compensation), automatic dilution in testo 320-2 LL Technical Data
-
Measuring range 0 to 15000 ppm
-
Accuracy ±200 ppm or ±20 % of mv (0 to 15000 ppm)
-
Resolution 1 ppm
Flue gas NO Technical Data
-
Measuring range 0 to 3000 ppm
-
Accuracy ±5 ppm (0 to 100 ppm)
-
±5 % of mv (101 to 2000 ppm)
-
±10 % of mv (2001 to 3000 ppm)
-
Resolution 1 ppm
Flue gas Draught Technical Data
-
Measuring range -9.99 to +40 hPa
-
Accuracy ±0.02 hPa or ±5 % of mv (-0.50 to +0.60 hPa
-
±0.03 hPa (+0.61 to +3.00 hPa)
-
±1.5 % of mv (+3.01 to +40.00 hPa)
-
Resolution 0.01 hPa
Temperature Technical Data
-
Measuring range -40 to +1200 °C
-
Accuracy ±0.5 °C (0.0 to +100.0 °C)
-
±0.5 % of mv (Remaining Range)
-
Resolution 0.1 °C (-40 t1 +999.9 °C)
-
1 °C (Remaining Range)
-
dependent on the thermocouple
Flue gas degree of effectivity, Eta (calculated) Technical Data
-
Measuring range 0 to 120 %
-
Resolution 0.1 %
Flue gas loss (calculated) Technical Data
-
Measuring range 0 to 99.9 %
-
Resolution 0.1 %
Flue gas CO₂ calculation (calculated from O₂) Technical Data
-
Measuring range Display range 0 to CO₂ max
-
Accuracy ±0.2 Vol.%
-
Resolution 0.1 Vol.%
Flue gas CO (without H₂-compensation) Technical Data
-
Measuring range 0 to 4000 ppm
-
Accuracy ±20 ppm or ±10 % of mv (0 to 400 ppm)
-
±5 % of mv (401 to 2000 ppm)
-
±10 % of mv (2001 to 4000 ppm)
-
Resolution 1 ppm
General Technical Data
-
Dimensions 270 x 160 x 57 mm
-
Power supply Rech. batt. block 3.7 V / 2.6 Ah; Mains unit 6 V / 1.2 A (optional)
-
System requirements requires iOS 7.1 or newer; requires Android 4.3 or newer; requires mobile end device with Bluetooth 4.0
-
Maximum 500000 readings
-
Operating temperature -5 to +45 °C
-
Storage temperature -20 to +50 °C
-
Weight 720 g (excluding battery)