Differential Pressure - Piezoresistive
-
Measuring range ±10000 Pa
-
Accuracy ±0.3 Pa (0 to 9.99 Pa) plus ±1 Digit
-
±3 % of mv (10 to 10000 Pa) plus ±1 Digit
Flue gas O₂
-
Measuring range 0 to 21 Vol.%
-
Accuracy ±0.2 Vol.%
-
Resolution 0.1 Vol.%
-
Reaction time t₉₀ < 20 s
Flue gas CO (with H₂-compensation)
-
Measuring range 0 to 8000 ppm
-
Accuracy ±10 ppm or ±10 % of mv (0 to 200 ppm)
-
±20 ppm or ±5 % of mv (201 to 2000 ppm)
-
±10 % of mv (2001 to 8000 ppm)
-
Resolution 1 ppm
-
Reaction time t₉₀ < 60 s
-
Measuring rate 0 to 500 ppm
Flue gas COlow
-
Measuring range 0 to 500 ppm
-
Accuracy ±2 ppm (0 to 39.9 ppm)
-
±5 % of mv (40 to 500 ppm)
-
Resolution 0.1 ppm
-
Reaction time t₉₀ < 40 s
Flue gas NO
-
Measuring range 0 to 3000 ppm
-
Accuracy ±5 ppm (0 to 100 ppm)
-
±5 % of mv (101 to 2000 ppm)
-
±10 % of mv (2001 to 3000 ppm)
-
Resolution 1 ppm
-
Reaction time t₉₀ < 30 s
Flue gas Nolow
-
Measuring range 0 to 300 ppm
-
Accuracy ±2 ppm (0 to 39.9 ppm)
-
±5 % of mv (40 to 300 ppm)
-
Resolution 0.1 ppm
-
Reaction time t₉₀ < 30 s
Flue Gas Draught
-
Measuring range -9.99 to +40 hPa
-
Accuracy ±0.02 hPa or ±5 % of mv (-0.50 to +0.60 hPa)
-
±0.03 hPa (+0.61 to +3.00 hPa)
-
±1.5 % of mv (+3.01 to +40.00 hPa)
-
Resolution 0.01 hPa
Temperature
-
Measuring range -40 to +1200 °C
-
Accuracy ±0.5 °C (0 to +100.0 °C)
-
±0.5 % of mv (Remaining Range)
-
Resolution 0.1 °C (-40 to +999.9 °C) ,1 °C (> +1000 °C)
Flue gas degree of effectivity, Eta (calculated)
-
Measuring range 0 to 120 %
-
Resolution 0.1 %
Flue gas loss (calculated)
-
Measuring range 0 to 99.9 %
-
Resolution 0.1 %
Flue gas CO₂ calculation (calculated from O₂)
-
Measuring range 0 to CO₂ max (Display range)
-
Accuracy ±0.2 Vol.%
-
Resolution 0.1 Vol.%
-
Reaction time t₉₀ < 40 s
Velocity / Volume flow
-
Measuring range 0.15 to 3 m/s
-
Resolution 0.1 m/s
Pressure measurement
-
Measuring range 0 to +300 hPa
-
Accuracy ±0.5 hPa (0.0 to 50.0 hPa)
-
±1 % of mv (50.1 to 100.0 hPa)
-
±1.5 % of mv (Remaining Range)
-
Resolution 0.1 hPa
Flue gas CO (without H2-compensation)
-
Measuring range 0 to 4000 ppm
-
Accuracy ±20 ppm (0 to 400 ppm)
-
±5 % of mv (401 to 2000 ppm)
-
±10 % of mv (2001 to 4000 ppm)
-
Resolution 1 ppm
-
Reaction time t₉₀ < 60 s
Ambient CO
-
Measuring range 0 to 500 ppm
-
Accuracy ±5 ppm (0 to 100 ppm)
-
±5 % of mv (> 100 ppm)
-
Resolution 1 ppm
-
Reaction time Approx. 35 s
-
with CO probe
Ambient CO₂
-
Measuring range 0 to 1 Vol.% , 0 to 10000 ppm
-
Accuracy ±50 ppm or ±2 % of mv (0 to 5000 ppm)
-
±100 ppm or ±3 % of mv (5001 to 10000 ppm)
-
Reaction time Approx. 35 s
-
with ambient CO₂ probe
Gas leak measurement for combustible gases (via gas leak detection probe)
-
Measuring range 0 to 10000 ppm CH₄ / C₃H₈; Display range
-
Accuracy Signal optical display (LED) audible signal via buzzer
-
Reaction time t₉₀ < 2 s
-
with gas leak detection probe
Temperature (via the fine pressure probe)
-
Measuring range -40 to +1200 °C max. (dependent on probe)
-
Accuracy ±0.5 °C (-40 to 100 °C)
-
±0.5 % of mv (Remaining Range) plus probe accuracy
-
Resolution 0.1 °C
General Technical Data
-
Protection class IP40
-
Dimensions 270 x 90 x 65 mm
-
Display size 240 x 320 pixels
-
Display funktion Colour graphic display
-
Power supply Rechargeable battery pack 3.7 V / 2.6 Ah; Mains unit 6 V / 1.2 A
-
Maximum 500,000 readings
-
Operating temperature -5 to +45 °C
-
Storage temperature -20 to +50 °C
-
Weight 600 g (without rechargeable battery)