Temperature - Type K TC Technical Data
-
Measuring range -4° to 212 °F / -20 to +100 °C
-
Accuracy ±2 °F / ±1 °C
-
Resolution 0.1 °F / 0.1 °C
-
Reaction time < 50 sec
-
Temperature (ambient temperature)
Temperature - Type J TC Technical Data
-
Measuring range 32° to 752 °F / 0 to +400 °C
-
Accuracy ±1.8 °F (32° to 212 °F) / ±1 °C (0 to +100 °C)
-
±1.5 % of remaining measurement value (> 212 °F) / ±1.5 % of mv (> 100 °C)
-
Resolution 0.1 °F / 0.1 °C
-
Reaction time < 50 sec
-
Temperature (flue gas)
Flue gas O₂ Technical Data
-
Measuring range 0 to 21 Vol.%
-
Accuracy ±0.2 Vol.%
-
Resolution 0.1 Vol.%
-
Reaction time t₉₀ 30 sec
Flue Gas Draft Technical Data
-
Measuring range -20 to +20 hPa
-
Accuracy ±0.03 hPa (-3.00 to +3.00 hPa)
-
±1.5 % of remaining measurement value (Remaining Range)
-
Resolution 0.01 hPa
Efficiency (calculated) Technical Data
-
Measuring range 0 to 120 %
-
Resolution 0.1 %
Flue Gas loss (calculated) Technical Data
-
Measuring range 0 to 99.9 %
-
Resolution 0.1 %
Flue Gas CO₂ calculation (calculated from O₂) Technical Data
-
Measuring range 0 to CO₂ max (Display range)
-
Accuracy ±0.2 Vol.%
-
Resolution 0.1 Vol.%
-
Reaction time t₉₀ < 40 sec
Pressure Measurement Technical Data
-
Measuring range -40 to +40 hPa
-
Accuracy ±0.5 hPa
-
Resolution 0.1 hPa
Flue gas CO (without H₂-compensation) Technical Data
-
Measuring range 0 to 4000 ppm
-
Accuracy ±20 ppm (0 to 400 ppm)
-
±5 % of remaining measurement value (401 to 2000 ppm)
-
±10 % of remaining measurement value (2001 to 4000 ppm)
-
Resolution 1 ppm
-
Reaction time t₉₀ 60 sec
Ambient CO Technical Data
-
Measuring range 0 to 4000 ppm
-
Accuracy ±20 ppm (0 to 400 ppm)
-
±5 % of remaining measurement value (401 to 2000 ppm)
-
±10 % of remaining measurement value (2001 to 4000 ppm)
-
Resolution 1 ppm
-
Reaction time 60 sec
General Technical Data
-
Dimensions 8 x 3 x 2 in. / 201 x 83 x 44 mm
-
Display type LCD
-
Display function Backlit 2-line display
-
Power supply Battery: 1500 mAh, mains unit 5V/1A
-
Maximum No memory
-
Operating temperature 23° to 113 °F / -5 to +45 °C
-
Storage temperature -4° to 122 °F / -20 to +50 °C
-
Weight (with probe) Approx. 25 oz. / (with probe) Approx. 700 g