Temperature - TC Type K (NiCr-Ni)
-
Measuring range: -40 to +1200 °C
-
Accuracy:
-
±0.5 °C (0.0 to +100.0 °C)
-
±0.5 % of mv (Remaining Range)
-
Resolution:
-
0.1 °C (-40 to +999.9 °C)
-
1 °C (Remaining Range)
Differential Pressure - Piezoresistive
-
Measuring range: -100 to +200 hPa
-
Accuracy:
-
±0.5 hPa (0 to +50.0 hPa)
-
±1 % of mv (+50.1 to +100.0 hPa)
-
±1.5 % of mv (+100.1 to +200 hPa)
-
Resolution: 0.01 hPa
Flue Gas O₂
-
Measuring range: 0 to 21 Vol.%
-
Accuracy: ±0.2 Vol.%
-
Resolution: 0.1 Vol.%
Differential Pressure Flue Gas Draught - Piezoresistive
-
Measuring range: -9.99 to +40 hPa
-
Accuracy:
-
±0.005 hPa (+0 to +0.1 hPa)
-
±0.02 hPa (+0.1 to +3.00 hPa)
-
±1.5 % of mv (3.01 to +40 hPa)
-
Resolution:
-
0.001 hPa (0 to 0.1 hPa)
-
0.01 hPa (Remaining Range)
Flue Gas Degree of Effectivity, ETA (calculated)
-
Measuring range: 0 to 120 %
-
Resolution: 0.1 %
Flue Gas Loss (calculated)
-
Measuring range: 0 to 99.9 %
-
Resolution: 0.1 %
Flue gas CO₂ calculation (calculated from O₂)
-
Measuring range: Display range 0 to CO₂ max
-
Accuracy: ±0.2 Vol.%
-
Resolution: 0.1 Vol.%
Flue gas CO (without H₂-compensation)
-
Measuring range: 0 to 4000 ppm
-
Accuracy:
-
±20 ppm (0 to 400 ppm)
-
±5 % of mv (401 to 2000 ppm)
-
±10 % of mv (2001 to 4000 ppm)
-
Resolution: 1 ppm
Ambient CO
-
Measuring range: 0 to 2000 ppm
-
Accuracy:
-
±10 ppm (0 to 100 ppm)
-
±10 % of mv (101 to 2000 ppm)
-
Resolution: 1 ppm
General Technical Data
-
Operating humidity: 15 to 90 %RH non-condensing
-
Dimensions: 244 (including probe connection) x 98 x 59 LxWxH
-
Operating temperature: -5 to +45 °C
-
Protection class: IP40
-
Product colour: Black/orange
-
Display type: 5.0" touch display, HD 1280*720 pixels, IPS (160 k)
-
Power supply: Rechargeable battery, USB mains unit
-
Guarantee: Gas sensors (O₂, CO) 24 months; rechargeable battery: 12 months; Thermocouple 12 months
-
Authorizations: TÜV-tested according to 1st German Federal Immission Control Ordinance (BImSchV) EN 50379, Parts 1-3
-
Warranty conditions: https://www.testo.com/guarantee
-
Memory: 1,000,000 measuring values
-
Storage battery type: Lithium ion battery
-
Battery life: 10 h
-
Storage temperature: -20 to +50 °C
-
Weight: approximately 800 g
-
Charging temperature: 0 to +45 °C
Differential Pressure 4 Pa measurement - Piezoresistive
-
Measuring Range: -50 to +50 Pa
-
Accuracy:
-
±0.3 Pa (< 10 Pa)
-
±3 % of mv (Remaining Range)
-
Resolution: 0.1 Pa